Certificate for ATEX Heated Hoses and ATEX Heated Lines
|
|
|
|

Certified Quality Management ISO 9001:2008 & Directive 94/9/EC Annex VII (ATEX)
|
|
|
|
|
Winkler Industrial Heating Mantles
are employed in the pharmaceutical, chemical and metal recycling industry in order to support chemical and biological reactions or distillation processes. The heat input is ensured by a heat conductor which is crocheted into the carrier material.
The basic design of this heating mantle series corresponds to the PILZ laboratory heating mantles. However, the industrial mantles are mechanically reinforced because of the flask sizes. The mounting fixture includes four sturdy eyebolts, loosely connected to a support ring.
The safety heating mantle is equipped with an additional temperature sensor (300 °C) per heating zone. There are suitable control devices available for both versions.
Advantages
- Hand-made products
- Even heat distribution up to a glass flask volume of 200 litres
- Built-in temperature sensor
- 3 to 5 heating zones (depending on size)
- Excellent workmanship combined with high-quality materials to guarantee a long service life
|
|
| Model WHI Industrial Heating Mantle |
Tmax = 450°C |
Description
- Outer jacket made of PTFE or aluminium coated
glass fabric (as desired)
- Hand-crocheted heating unit made of glass silk
- 3 to 5 heating zones
- Suitable for flasks from 10 l up to 200 l
Device features
- Terminal box
- Connecting cable with 10-pole industrial plug (2.0 m)

|
|
|
| Model WHIS Safety Heating Mantle |
Tmax = 300°C |
Description
- Outer jacket made of PTFE or aluminium coated glass fabric (as desired)
- Hand crocheted heating unit made of glass silk
- 3 to 5 heating zones
- Suitable for flasks from 10 l up to 200 l
- Combined with a Winkler WRTI72 controller
for temperature control to a maximum of 300 °C
- Protects the glass flasks against stress. In the event of excessive stress
there is a danger of glass fl ask breakage.
Device features
- Terminal box
- Connecting cable with 10-pole industrial plug (2.0 m)
- cable with sensor plug (2.0 m)
- one temperature sensor per heating zone
- Standard: Thermo element Type K (NiCr-Ni) Option: Thermo element Type J (Fe-CuNi)
Options: Thermocouple Type K (NiCr-Ni), thermocouple Type J (Fe-CuNi), RTD Pt 100

|
|
|
| Controller WRTI71 and WRTI72 |
Description and device features (Controller WRTI71 and WRTI72)
- Table-top controller in a powder coated, robust metal housing
with high chemical resistance
- Aluminium front plate
- Master switch
- Digital temperature display for lowest heating zone
- 3 to 5 power selectors with control status display for each heating zone
- With CEE-plug 16A for three-phase current connection 3×230V/50Hz
(cable length 1.5 m)
- Multipole, plug-in device outlet for industrial heating mantle connection
Controller WRTI Controller series WRTI71 is employed for operating industrial heating mantles.

|
|
|
| Controller WRTI72 for Safety Heating Mantles |
|
Controller series WRTI72 incorporates a safety system that limits the temperature of the different heating zones to 300 °C.
Additional features of WRTI72 device
- 10-pole sensor connection
- Cable entry for signal contact to permit temperature monitoring
- Lemo socket size 1; 4-pole; assignment in accordance with NAMUR
(connection for external sensor to measure the media temperature)
- The overall control takes place via the freely selectable medium temperature Tmax=300 °C
- The temperature is limited by built-in temperature controllers -> 300 °C for each zone being heated
|
|
|
|
|
|